Description | L1B Impact | L2 Impact | Cause | Mitigation | Schedule | History DESCRIPTION:Band 21 calibration is imprecise at high temperatures. IMPACT TO L1B DATA SET:This effect causes increased uncertainty in the brightness temperature due to the indirect nature of the calibration. Detector to detector striping exists, but is within specifications. ESTIMATED IMPACT TO L2 SCIENCE PRODUCTS:Science research will be seriously impacted until the calibration can be extended. CAUSE OF EFFECT:The saturation brightness temperature for this fire detection band is specified at 500 K. Pre-launch test targets had a maximum brightness temperature of 340 K, which equals about 3.5% of full radiance scale. Band 21 is expected exhibit the non-linear characteristics seen in other MWIR bands, but the calibrated temperature range is not adequate for determining non-linear calibration characteristics for this band. MITIGATION APPROACH/ADDED TESTING TO CHARACTERIZE EFFECT: The linear gain coefficient for
this band is computed nearly monthly using an elevated temperature
cycle of the OBC-BB. LUTS are updated as needed. ESTIMATED SCHEDULE TO COMPLETE:TBD. Gains are updated as needed. HISTORY OF CHARACTERISTIC: